The polymer-based SEC (GFC) columns with additional separation mode

Product information

Multi mode chromatography is an analytical method combining several separation modes. 

When a single separation mode is not enough for separating the target substance, an improvement can be expected by using multi mode chromatography. The balance between the separation modes involved can be controlled by changing the eluent composition.

Asahipak GS series [multi-mode columns]

  • Polymer-based columns

  • SEC is the main separation mode

  • With the choice of eluent, the column provides multimode features of reversed phase, HILIC, and ion exchange modes to SEC

  • Suitable for the separation of peptides or nucleic acids with similar molecular weights

  • Suitable for desalting samples or substituting buffer in protein analysis

The base material is a polymer, resulting in the following advantages;

1) usable at a wide pH range, 2) washable with alkaline solutions, 3) column durability.

Asahipak GS columns operate basically in SEC mode. However, depending on the eluent conditions, it allows separation by multiple modes combining reversed phase and ion exchange modes. The column is particularly effective in analyzing samples such as proteins and nucleic acid components, which are made up of both acidic and basic components. In this separation mode, samples are eluted in the order of acidic, neutral and basic. Separation can also be adjusted by changing the eluent components.

1) When a neutral/acidic eluent is used.

Acidic samples: As it is the ion exclusion mode which is in operation, substances are eluted relatively quickly.

Neutral samples: As substances are separated by reversed phase mode, they are eluted moderately.

Basic samples: As it is the ion exchange mode which is in operation, substances are eluted relatively slow.


2) When a basic eluent is used.

Acidic samples: Basically, substances are excluded together at the V(0) position but, with acidic samples of large molecular weights, there are cases in which substances are retained in the column.

Neutral samples: Substances are separated by reversed phase mode.

Basic samples: If the eluent has a higher pH than the pKa of the sample, substances are separated by reversed phase mode as the dissociation of basic samples is suppressed.

Product overview

Analytical Columns

Product code
Product name
Separation
Functional group
Gel material
Plates per column
Particle size
Pore size
Size (ID x length)
Housing material
F6710019
Asahipak GS-2G 7B
SEC
-
Polyvinyl alcohol
(guard)
9 µm
-
7.5 x 50 mm
steel
F7600006
Asahipak GS-320 HQ
SEC
-
Polyvinyl alcohol
≥ 19,000
6 µm
400 Å
7.5 x 300 mm
steel
F7600005
Asahipak GS-220 HQ
SEC
-
Polyvinyl alcohol
≥ 19,000
6 µm
150 Å
7.5 x 300 mm
steel

Preparative Columns

Product code
Product name
Separation
Functional group
Gel material
Plates per column
Particle size
Pore size
Size (ID x length)
Housing material
F6810034
Asahipak GS-220 20G
SEC
-
Polyvinyl alcohol
≥ 14,000
13 µm
150 Å
20.0 x 500 mm
steel
F6810035
Asahipak GS-320 20G
SEC
-
Polyvinyl alcohol
≥ 14,000
13 µm
400 Å
20.0 x 500 mm
steel

Product details

Analytical Columns

Product name
Maximum pressure
Usual flow rate
Maximum flow rate
Temperature range
pH range
Salt concentration
Shipping solvent
USP
MW range
Exclusion limit
Asahipak GS-2G 7B
-
-
-
4 to 60°C
2 to 12
≤ 0.5 M
H2O/CH3OH = 70/30
-
-
-
Asahipak GS-320 HQ
5.0 MPa (50 bar)
0.4 to 0.6 mL/min
1.0 mL/min
4 to 60°C
2 to 12
≤ 0.5 M
H2O/CH3OH = 70/30
-
300 - 20,000
40,000
Asahipak GS-220 HQ
6.0 MPa (60 bar)
0.4 to 0.6 mL/min
1.0 mL/min
4 to 60°C
2 to 9
≤ 0.5 M
H2O/CH3OH = 70/30
-
300 - 3,000
7,000

Preparative Columns

Product name
Maximum pressure
Usual flow rate
Maximum flow rate
Temperature range
pH range
Salt concentration
Shipping solvent
USP
MW range
Exclusion limit
Asahipak GS-220 20G
-
4.0 to 6.0 mL/min
8.0 mL/min
4 to 60°C
2 to 9
≤ 0.5 M
H2O/CH3OH = 70/30
-
300 - 3,000
7,000
Asahipak GS-320 20G
-
5.0 to 8.0 mL/min
12 mL/min
4 to 60°C
2 to 12
≤ 0.5 M
H2O/CH3OH = 70/30
-
300 - 20,000
40,000

Product info

Applications

At the moment a new window opens and you are transferred to our Japanese website, when you click on the application. We are working on a better solution.

Calibration Curves for GS-HQ Series (Aqueous Eluent: PEG, PEO)

Calibration Curves for GS-HQ Series (Aqueous Eluent: Pullulan)

Eluent Conditions for Asahipak GF, GS HQ series

Arsenic Compounds in Marine Life

Cyclodextrins

Elution Volume of Saccharides

Hydrophobicity of Columns

Low-Molecular-Weight Water Soluble Dietary Fiber (GS-220 HQ, KS-802)

Phytic Acid (GS-220 HQ)

Poly(Ethylene Glycol) Standards (9) (GS-220 HQ)

Polyphosphates (3) (GS-220 HQ)

Polyphosphates (4) (GS-220 HQ)

Pullulan Standards (7) (GS-220 HQ)

"Umami" (GS-320 HQ)

Adenine Nucleotides in Red Blood Cell

Adenosine and Derivatives

Analysis of ATP and its Related Substances in Muscle of Fish and Shellfish (GS-320 HQ)

Analysis of Peptide using Multimode Column

Analysis of Uric Acid and Creatinine using Column Switching Method

Analysis of VMA and Creatinine using Column Switching Method

Ascorbic Acid and Uric Acid in Blood

Ascorbic Acid and Uric Acid in Liquor Cerebronspinalis

Ascorbic Acid in Urine

Chymotrypsin-Hydrolysated Peptides of Ascidian 29K Protein

Crude Pronase E

Cyclic AMP and AMP

Effects of Analysis of PEG Mixtures with GS-320 HQ by Composition of Eluent

Endorphins

Glutathione in Blood

HVA and VMA in Urine

Hydrophobicity of Columns

Hydroxyriboflavin and Riboflavin in Urine

Influence of Column Temperature on Retention Time of PEG (GF-310 HQ and GS-320 HQ)

Isolation of Ecdyson from Vitex Fiherii

Methamphetamine in Urine

Nucleobases (1) (GS-320 HQ)

Nucleobases, Nucleosides and Nucleotides (GS-320 HQ)

Nucleosides (1) (GS-320 HQ)

Oligo-DNA (GS-320 7G)

Paraquat in Rat Serum

Peptides in Extract from Waste Water Sludge

Preparation of CS-Glucan in Coix Seed

Purine Base in Beer (GS-320 HQ)

Purine Bases and Uric Acid Separation (GS-320 HQ)

Recovery of Proteins (GF-310 HQ and GS-320 HQ)

Riboflavin (Vitamin B2)

Standard Proteins (12) (GS-320 7G)

Theophylline in Serum

Related products

Shodex STANDARD P-82 (Pullulan) is a calibration standard kit for aqueous size exclusion chromatography.

Pullulan is an unbranched polysaccharide: units of maltotriose are connected with α-1,6 bond.

  • Easy to prepare

  • For aqueous SEC (GFC)

  • Unbranched pullulan standard

  • High solubility in water eliminates the possibility of recrystallization

STANDARD P-82 kit consists of 8 vials with different molecular weight standards of 200 mg each:

MW 800.000, 400.000, 200.000, 100.000, 50.000, 20.000, 10.000, 5.000​