top of page
Shodex PROTEIN Series_43.jpg

PROTEIN KW/ PROTEIN LW columns

separation by size also for high molecular weight

The silica-based SEC (GFC) column for proteins and enzymes

Product information

PROTEIN KW-800 series
The packing material of the Shodex PROTEIN KW-800 series is based on silica particles whose surface is coated with a hydrophilic polymer. This column is for aqueous SEC (Gel filtration chromatography (GFC)). It enjoys a positive reputation thanks to its high recovery ratio of proteins and peptide. The PROTEIN KW-800 series comprises three kinds of columns, differing in pore and particle size.

  • Silica-based packed columns for aqueous SEC (GFC) analysis

  • Suitable for the analysis of proteins and enzymes

  • Fulfills USP-NF L20, L33, and L59 requirements

PROTEIN KW400 series
The Shodex KW400-4F series is the high-performance semi-micro column version of the KW-800 series. The theoretical plate number and the sensitivity are improved by reducing the diameter of the packed particle to 3 micrometers. KW405-4F, the newest addition to the KW400 line-up, has a larger pore size than any columns in the existing KW-800 series, and can analyze proteins whose molecular weight is larger than 1,000,000. It enables the scientist to carry out the analysis of protein complexes.

  • Reduced packing material particle size enhances column performance

  • Three to four-fold higher sensitivity than KW-800 series

  • KW405-4F is applicable analyzing samples with molecular weight above 1,000,000

  • Fulfills USP-NF L20, L33, and L59 requirements

PROTEIN LW-803 and LW-403 4D

More and more monoclonal antibodies (mAb) are used for therapeutic drugs. The pore size of the Shodex PROTEIN LW-803 and LW-403 4D is suitable for this analysis.

  • Pore size specifically controlled for analyzing proteins with a molecular weight of several hundred of thousand

  • High performance analysis of antibody drugs and various proteins

  • High lot-to-lot reproducibility

  • Fulfills USP-NF L20, L33, and L59 requirements

Product overview

Product code
Product name
Separation
Functional group
Gel material
Plates per column
Particle size
Pore size
Size (ID x length)
Housing material
F6989000
PROTEIN KW-802.5
SEC
Hydroxyl
Silica
≥ 21,000
5 µm
150 Å (max. 400 Å)
8.0 x 300 mm
steel
F6989103
PROTEIN KW-803
SEC
Hydroxyl
Silica
≥ 21,000
5 µm
300 Å (max. 1,000 Å)
8.0 x 300 mm
steel
F6989104
PROTEIN KW-804
SEC
Hydroxyl
Silica
≥ 16,000
7 µm
500 Å (max. 1,500 Å)
8.0 x 300 mm
steel
F6700131
PROTEIN KW-G 6B
SEC
Hydroxyl
Silica
(guard)
7 µm
-
6.0 x 50 mm
steel
F6989201
PROTEIN KW402.5-4F
SEC
Hydroxyl
Silica
≥ 35,000
3 µm
150 Å (max. 400 Å)
4.6 x 300 mm
steel
F6989202
PROTEIN KW403-4F
SEC
Hydroxyl
Silica
≥ 35,000
3 µm
250 Å (max. 800 Å)
4.6 x 300 mm
steel
F6989203
PROTEIN KW404-4F
SEC
Hydroxyl
Silica
≥ 25,000
5 µm
500 Å (max. 1,500 Å)
4.6 x 300 mm
steel
F6989204
PROTEIN KW405-4F
SEC
Hydroxyl
Silica
≥ 25,000
5 µm
1,000 Å (max. 2,000 Å)
4.6 x 300 mm
steel
F6700132
PROTEIN KW400G-4A
SEC
Hydroxyl
Silica
(guard)
5 µm
-
4.6 x 10 mm
steel
F6989303
PROTEIN LW-803
SEC
Hydroxyl
Silica
≥ 12,000
3 µm
300 Å (max. 1,000 Å)
8.0 x 300 mm
steel
F6700133
PROTEIN LW-G 6B
SEC
Hydroxyl
Silica
(guard)
3 µm
-
6.0 x 50 mm
steel
F6989403
PROTEIN LW-403 4D
SEC
Hydroxyl
Silica
≥ 11,000
1.9 µm
300 Å (max. 1,000 Å)
4.6 x 150 mm
steel
F6700134
PROTEIN LS-G 4J
SEC
Hydroxyl
Silica
(guard)
1.9 µm
-
4.6 x 20 mm
steel

Product details

Product name
Maximum pressure
Usual flow rate
Maximum flow rate
Temperature range
pH range
Salt concentration
Shipping solvent
USP
MW range
Exclusion limit
PROTEIN KW-802.5
5.0 MPa (50 bar)
1.0 mL/min
1.5 mL/min
4 to 45°C
3.0 to 7.5
≤ 0.3 M
H2O
L20, L33, L59
5,000 - 100,000
150,000
PROTEIN KW-803
5.0 MPa (50 bar)
1.0 mL/min
1.5 mL/min
4 to 45°C
3.0 to 7.5
≤ 0.3 M
H2O
L20, L33, L59
10,000 - 700,000
(1,000,000)
PROTEIN KW-804
5.0 MPa (50 bar)
1.0 mL/min
1.5 mL/min
4 to 45°C
3.0 to 7.5
≤ 0.3 M
H2O
L20, L33, L59
30,000 - (4,000,000)
(4,000,000)
PROTEIN KW-G 6B
-
-
-
4 to 45°C
3.0 to 7.5
≤ 0.3 M
H2O
L20, L33, L59
-
-
PROTEIN KW402.5-4F
10 MPa (100 bar)
0.35 mL/min
0.5 mL/min
5 to 45°C
3.0 to 7.5
≤ 0.3 M
H2O
L20, L33, L59
5,000 - 70,000
150,000
PROTEIN KW403-4F
10 MPa (100 bar)
0.35 mL/min
0.5 mL/min
5 to 45°C
3.0 to 7.5
≤ 0.3 M
H2O
L20, L33, L59
10,000 - 500,000
600,000
PROTEIN KW404-4F
7.0 MPa (70 bar)
0.35 mL/min
0.5 mL/min
5 to 45°C
3.0 to 7.5
≤ 0.3 M
H2O
L20, L33, L59
30,000 - (4,000,000)
(4,000,000)
PROTEIN KW405-4F
7.0 MPa (70 bar)
0.35 mL/min
0.5 mL/min
5 to 45°C
3.0 to 7.5
≤ 0.3 M
H2O
L20, L33, L59
200,000 - (20,000,000)
(20,000,000)
PROTEIN KW400G-4A
-
-
-
5 to 45°C
3.0 to 7.5
≤ 0.3 M
H2O
L20, L33, L59
-
-
PROTEIN LW-803
15 MPa (150 bar)
1.0 mL/min
1.5 mL/min
4 to 45°C
3.0 to 7.5
≤ 0.3 M
H2O
L20, L33, L59
10,000 - 700,000
(1,000,000)
PROTEIN LW-G 6B
-
-
-
4 to 45°C
3.0 to 7.5
≤ 0.3 M
H2O
L20, L33, L59
-
-
PROTEIN LW-403 4D
20 MPa (200 bar)
0.35 mL/min
0.5 mL/min
5 to 45°C
3.0 to 7.5
≤ 0.3 M
H2O
L20, L33, L59
10,000 - 700,000
(1,000,000)
PROTEIN LS-G 4J
-
-
-
4 to 45°C
3.0 to 7.5
≤ 0.3 M
H2O
L20, L33, L59
-
-

Products

Product instruction

Usable solvents are listed below.
(1) The standard eluent is a buffer with a salt. Buffers such as phosphate, acetate, and Tris-HCl can be used.
(2) Salts such as sodium chloride, sodium sulfate, potassium sulfate and ammonium sulfate can be used. Please keep total concentration of salts under 1 M. Their recommended concentration ranges are 0.1 - 0.3 M.
(3) Up to 100 % (v/v) acetonitrile, methanol, ethanol, and 2-propanol (IPA) are usable.
(4) Protein denaturants such as urea and 6-M guanidine hydrochloride, and surfactants such as SDS and Brij-35 can also be used. However, since protein denaturants and surfactants tend to remain on the column, solvent replacement after their use takes a longer time than replacement from general solvents. It is recommended to dedicate the column for this specific use.

<Cleaning Method>
(1) Insoluble components that block the column inlet may be removed by reversing the flow direction, i.e., introducing the eluent from the column outlet, with flow rate at less than half of the recommended flow rate.
(2) Follow below cleaning steps for adsorbing components. For an efficient cleaning, reverse the flow direction and use the flow rate at less than half of the recommended flow rate. Introduce 5 to 10 times column volume of the washing solvent.
Method 1: Adsorption of ionic (basic) compounds
· Introduce the eluent with a salt concentration of 0.5 M.
· Introduce an acidic aqueous solution of pH 3.
Method 2: Adsorption of hydrophobic compounds
· Introduce the eluent with 10 - 20 % (v/v) organic solvent (acetonitrile or methanol).

Applications

At the moment a new window opens and you are transferred to our Japanese website, when you click on the application. We are working on a better solution.

General PROTEIN KW-800

Effects of Flow Rate (KW-802.5, KW-803)

Recovery of Proteins (KW-802.5 and KW-803)

Recovery of Proteins (KW-803)

Sample Load (KW-803)

Standard Proteins (1) (KW-802.5)

Standard Proteins (14) (KW-802.5)

Standard Proteins (15) (KW-803)

Standard Proteins (16) (KW-804)

Standard Proteins (18) (KW-803)

Standard Proteins (19) (KW-804)

Standard Proteins (2) (KW-802.5 and KW-803)

Calibration Curves for KW-800 Series (PEG, PEO)

Calibration Curves for KW-800 Series (Proteins)

Calibration Curves for KW-800 Series (Pullulan)

Analysis PROTEIN KW-800

α1 Acid Glycoprotein

Analysis of Enoxaparin Sodium (KW-803 + KW-802.5)

Analysis of Exenatide According to USP-NF Method (KW-802.5)

Analysis of High Molecular Impurities in Insulin Aspart According to Pharmacopeia Method (KW-802.5)

Analysis of High Molecular Weight Protein According to USP-NF GENERAL CHAPTERS Physicochemical Analytical Procedures for Insulins (KW-802.5)

Analysis of Insulin According to USP-NF Method (KW-802.5)

Analysis of Insulin Glargine According to JP Method (KW-802.5)

Analysis of Insulin Human According to JP Method (KW-802.5)

Analysis of Lipoproteins in Serum (KW-804)

Analysis of Semaglutide (KW-802.5)

Analysis of Teriparatide According to EP Method (KW-802.5)

Analysis of Thymosin α1 (KW-802.5)

Avidin (2)

BSA Analysis in Surfactant (On-line Surfactant Removal)

Comparison of Chromatogram of PEG Mixtures

Comparison of Chromatograms of Protein mixtures

Comparison of SEC Separation of Polyclonal Antibody

Comparison of SEC Separation of Standard Proteins

Comparison of Separation of BSA Aggregates

Crude Albumin (Chicken Egg) (1) (KW-803)

Crude Catalase

Crude Hexokinase

Crude Lipoxidase

Dimer and Trimer of BSA

Heparin (3) (KW-804)

PEG-Chymotrypsin (KW-802.5)

Poly-Lysine

Proteins in Human Serum (3) (KW-803)

SEC Analysis of Filgrastim (KW-803)

SEC Analysis of Liraglutide Formulation (KW-802.5)

Molecular Weight Distribution Analysis of Collagen Peptide According to Gelatin Manufactures Association of Japan Method (KW-802.5)

Molecular Weight Measurements of Low Molecular Weight Heparins (KW-803 + KW-802.5)

SEC Analysis of Etelcalcetide Hydrochloride (KW-802.5)

SEC Analysis of Low Molecular Weight Heparin (1) (Dalteparin Sodium) (KW-803 + KW-802.5)

SEC Analysis of Low Molecular Weight Heparin (2) (Parnaparin Sodium) (KW-803 + KW-802.5)

Tryptic Digest of Human IgG

General PROTEIN KW400

Comparison between KW402.5-4F and KW-802.5

Recovery of Proteins (KW402.5-4F, KW403-4F)

Standard Proteins (24) (KW402.5-4F)

Standard Proteins (25) (KW403-4F)

Standard Proteins (26) (KW404-4F)

Standard Proteins (27) (KW405-4F)

Calibration Curves of KW400-4F (1) (Pullulan)

Calibration Curves of KW400-4F (2) (Proteins)

Calibration Curves of KW400-4F (3) (PEG, PEO)

Analysis PROTEIN KW400

Analysis of Hemocyanin (KW405-4F)

Analysis of Whey in Yogurt (KW403 4F + KW402.5 4F)

Human Control Serum

Lectin Protein

Particle Size Distribution Determination of Hepatitis B Surface Antigen Virus-like Particle (HBsAg VLP) (KW405-4F)

Peptide Analysis (SEC)

SEC Analysis of Latex Particles (KW405-4F)

Separation of Ceftazidime and Arginine (KW402.5-4F)

General PROTEIN LW-803 and LW-403 4D

Calibration Curve for LW-803 (Proteins)

Calibration Curve for LW-403 4D (Proteins)

Analysis PROTEIN LW-803 and LW-403 4D

Analysis of Epoetin According to USP-NF Method (LW-803)

Analysis of Erythropoietin According to EP Method (LW-803)

Analysis of Somatropin According to EP Method (LW-803)

Comparison of SEC Separation of IgG between LW-403 4D and LW-803

Comparison of SEC Separation of Monoclonal Antibody

Comparison of SEC Separation of Polyclonal Antibody

Comparison of SEC Separation of Standard Proteins

Comparison of SEC Separation of Standard Proteins between LW-403 4D and LW-803

Comparison of Separation of BSA Aggregates

Monitoring Papain Digestion of Humanized Monoclonal IgG (LW-803)

SEC Analysis of Antibody Drug Conjugate (LW-803)

SEC Analysis of Rituximab (LW-803)

Comparison of IgG Separation with Conventional Device and Semi-micro Type Device (LW-403 4D)

Comparison of SEC Separation of IgG between LW-403 4D and LW-803

Comparison of SEC Separation of Standard Proteins between LW-403 4D and LW-803

Comparison of Separation of Proteins with Conventional Device and Semi-micro Type Device (LW-403 4D)

Durability Test by Number of Sample Injections (LW-403 4D)

Related products

Shodex STANDARD P-82 (Pullulan) is a calibration standard kit for aqueous size exclusion chromatography.

Pullulan is an unbranched polysaccharide: units of maltotriose are connected with α-1,6 bond.

  • Easy to prepare

  • For aqueous SEC (GFC)

  • Unbranched pullulan standard

  • High solubility in water eliminates the possibility of recrystallization

STANDARD P-82 kit consists of 8 vials with different molecular weight standards of 200 mg each:

MW 800.000, 400.000, 200.000, 100.000, 50.000, 20.000, 10.000, 5.000​